90
3-14-31. :TRIGger:TIMEOut:TIMER
Set
Query
Description
Sets or returns timeout trigger time.
Syntax
:TRIGger:TIMEOut:TIMER {<NRf> | ? }
Related
commands
:TRIGger:TIMEOut:WHEn
Parameter
<NRf>
Timeout time. (4nS to 10S).
Return parameter Returns the timeout time as <NR3>.
Example
:TRIGger:TIMEOut:TIMER?
8.960e-05
3-14-32. :TRIGger:ALTernate
Set
Query
Description
Sets alternating between source triggers on or off or
queries its state.
Syntax
:TRIGger:ALTernate {OFF | ON |?}
Parameter
OFF
Alternate off
ON
Alternate on
Return parameter
Returns the Alternate trigger status (ON, OFF).
Example
:TRIGger:ALTernate ON
Turns on alternating between source triggers.
3-14-33. :TRIGger:STATe
Query
Description
Returns the current state of the triggering system.
Syntax
:TRIGger:STATe?
Return parameter
*ARMED
Indicates that the oscilloscope is acquiring
pretrigger information.
*AUTO
Indicates that the oscilloscope is in the
automatic mode and acquires data even in
the absence of a trigger.
*READY
Indicates that all pretrigger information has
been acquired and that the oscilloscope is
ready to accept a trigger.
*SAVE
Indicates that the oscilloscope is in save
mode and is not acquiring data.
*TRIGGE
R
Indicates that the oscilloscope triggered
and is acquiring the post trigger
information.
Example
:TRIGger:STATe?
AUTO