69
3-10-2. :MEASUrement:MEAS<X>:TYPe
Set
Query
Description
Sets or queries the measurement type for a selected
automatic measurement. This is a statistics related
command.
Syntax
:MEASUrement:MEAS<X>:TYPe
{PK2pk | MAXimum | MINImum | AMPlitude | HIGH |
LOW | MEAN | CMEan | RMS | CRMs | AREa | CARea |
ROVShoot | FOVShoot | RPReshoot | FPReshoot |
FREQuency | PERIod | RISe | FALL | PWIdth | NWIdth |
PDUTy | PPULSE | NPULSE | PEDGE | NEDGE
| FRRDelay | FRFDelay | FFRDelay | FFFDelay |
LRRDelay | LRFDelay | LFRDelay | LFFDelay | PHAse
| ?}
Related commands :MEASUrement:MEAS<X>:SOURCE<X>
Parameter
MEAS<X> The automatic measurement number from
1 to 8.
Return parameter
Returns the measurement type
Example
:MEASUrement:MEAS1:TYPe RMS
Sets measurement 1 to RMS measurement.
3-10-3. :MEASUrement:MEAS<X>:STATE
Set
Query
Description
Sets or queries the state of a selected measurement.
This is a statistics related command.
Syntax
:MEASUrement:MEAS<X>:STATE { ON | OFF | 1 | 0
| ? }
Related commands :MEASUrement:MEAS<X>:SOUrce<X>
:MEASUrement:MEAS<X>:TYPe
Parameter
MEAS<X> The automatic measurement number from
1 to 8.
ON/1
Turn the measurement on.
OFF/0
Turn the measurement off.
Return parameter
0
Measurement is off.
1
Measurement is on.
Example
:MEASUrement:MEAS1:STATE 1
Turns measurement 1 on.