20
3-2-8. :ACQuire<X>:STATe?
Query
Description
Returns the status of waveform data.
Syntax
:ACQuire<X>:STATe?
Parameter
<X>
Channel number (1 to 4)
Return parameter
0
Raw data is not ready
1
Raw data is ready
Example
:ACQuire1:STATe?
0
Returns 0. The channel
1’s raw data is not ready.
Note: If the oscilloscope changes the acquisition status
from STOP to RUN, the status will be reset as zero.
3-2-9. :ACQuire:RECOrdlength
Set
Query
Description
Sets or queries the record length. Please see the user
manual for full details.
Syntax
:ACQuire:RECOrdlength {<NRf>| ?}
Parameter/Return
parameter
<NRf>
Record length. Settable record length: (1e+3 |
1e+4 | 1e+5 | 1e+6 | 1e+7)
Example
:ACQuire:RECOrdlength 1e+3
Sets the record length to 1000 points.