3-6-2. :CURSor:SOURce
Set
Query
Description
Sets or queries the cursor source.
Syntax
:CURSor:SOURce {CH1 | CH2 |CH3 | CH4 | REF1 |
REF2 | REF3 | REF4 | MATH | BUS1 | ?}
Parameter
CH1~CH4
Channel 1 to 4
REF1~4
Reference waveform 1 to 4
MATH
Math source
BUS1
Bus source
Return parameter Returns the cursor source.
Example
:CURSor:SOURce CH1
Turns the cursor source as channel 1.
3-6-3. :CURSor:HUNI
Set
Query
Description
Sets or queries the units for the horizontal bar cursors.
Syntax
:CURSor:HUNI {SEConds | HERtz | DEGrees |
PERcent | ?}
Related Commands :CURSor:MODe
Parameter
SEConds
Sets the cursor units to time in seconds.
HERtz
Sets the cursor units to frequency.
DEGrees
Sets the cursor units to degrees.
PERcent
Sets the cursor units to percent.
Return parameter
Returns the unit type.
Example
:CURSor:HUNI SEConds
Sets the units to time in seconds.
3-6-4. :CURSor:HUSE
Set
Description
Sets the current cursor position as the phase or ratio
reference for the Percent or Degrees (horizontal)
cursors.
Note
This command can only be used when
:CURSor:HUNI is set to DEGrees or PERcent.
Syntax
:CURSor:HUSE {CURRent}
Related Commands :CURSor:MODe
:CURSor:HUNI
Parameter
CURRent
Uses the current horizontal position
Example
:CURSor:HUSE CURRent.