5 Parameterizing, adjusting and analyzing software (P2A software)
72
- The values are reset to the customer-specific factory settings.
4.
Click on
[OK]
or
[Cancel]
to close the dialogue.
5.3.3.3.
Testing analog output
✓
The required instrument file is marked.
1.
Click on
[Test/analyze transmitter]
.
- The
Properties of <Instrument type> <Serial number>
dialogue is opened with the
Test/analyze transmitter
register.
2.
Mark channel and test values.
Field/button Explanation
Transmitter test
Check the analog outputs (see Testing
functionality of analog outputs, page
45).
Current reading
Readings are updated every second.
Unit
Unit according to the type of analog
output.
Default value
Freely definable output value for the
respective type of analog output (V or
mA), 1 decimal place.