5 Parameterizing, adjusting and analyzing software (P2A software)
69
Field
Explanation
The accuracy specifications are only valid for the zeroing
cycle of 15 sec set by the factory.
Zeroing cycle setting Selection of the time interval after which the
automatic zeroing cycle is to occur.
Recommendation: 15 sec
5.3.2.2.
Saving parameters
Parameters can be saved in new parameter files.
1.
Mark instrument/parameter file.
2.
Click on
File > Save as
in the menu bar.
3.
Select storage location and enter the file name.
4.
Click on
[Save]
.
- The new parameter file is shown in the file list.
Only the parameters are saved from an instrument file; the history
data are not adopted.
The original name (Instrument type, Serial number) is
suggested with the current date/time as standard, e.g.
"testo 6383 01234578 061120 1403.cfp".
For a standard installation, the files are saved under
"C:\Documents and Settings\All Users\Shared
Documents\P2A Software". The path can differ depending
on the version of the operating system.