4 Transmitter
24
4.3.4.
Ethernet communication
4.3.4.1.
Types of operation
The type of operation can be set via the operating menu:
Settings
|
Ethernet
|
Mode
|
•
Saveris Probe Mode
: a corresponding testo 6383 becomes a
Saveris participant
•
XML Server Mode
: a corresponding testo 6383 can be
integrated into unlimited Ethernet systems
4.3.4.2.
Mains connection
> Connect the network cable to the transmitter.
The network cable must not be connected directly to the
telephone network (ISDN).
4.3.4.3.
testo as a Saveris component
Set IP address
✓
Network cable must be connected to the Ethernet interface.
✓
Service plug must be connected.
The IP address of the transmitter can set be either via the P2A
software (transmitter accessory) or via the configuration menu of
testo Saveris
TM
(testo Saveris
TM
Ethernet Assistent).
1.
Set the IP address with the P2A software (see
Using the
software
, page 59) or with the testo Saveris
TM
Ethernet
Assistent (see instruction manual).
2.
Remove service plug.
Using Saveris
> Start
Saveris software (see instruction manual measurement
data monitoring with testo Saveris).
Ensure that the Saveris software is not already open, for
example in multi-user operation in Windows
®
Vista. If
several clients are installed in one network, ensure that
when clients are operating simultaneously, they do not
carry out simultaneous alterations to the configuration of
the system.