4 Transmitter
81
Message Display
Description
02102
2-point adjustment
11.3 %
As part of the 2-point adjustment,
an adjustment is performed at
11.3 % RH
02103
2-point adjustment
75.3 %
As part of the 2-point adjustment,
an adjustment is performed at
75.3 % RH
02105
Self-adjustment
active
For testo 6615 probe only: The
probe performs an automatic self-
adjustment
Pos: 82 /TD/Produkt verwenden/MUF 63xx/Status-Warnmeldungen/Warnmeldungen 638x @ 3\mod_1234534609966_79.doc @ 25493 @ 3
4.5.2. Warning
messages
Warning messages show an early warning or a current malfunction
which may negatively impact measuring.
Message Display
Cause
Remedying of fault
00809
Pressure too
high**
The process pressure
exceeds the pressure
intended for the
transmitter
Remove the transmitter
from the process and
take any necessary
measures to lower the
pressure
00E00
T ambient high** The ambient
temperature exceeds
the permissible
temperature for the
transmitter
Take necessary
measures to lower
ambient temperature,
e.g. through venting or
cooling
00E01
T ambient low** The ambient
temperature is below
the permissible
temperature for the
transmitter
Take necessary
measures to raise
ambient temperature,
e.g. through heating
00E04/00E
05
Supply voltage
low**
The supply voltage is
below the minimum
voltage required for the
transmitter
Take measures to
ensure sufficient
voltage supply
02822
T process high** The process
temperature exceeds
the temperature
designated for the
probe
Remove the probe from
the process and take
any necessary
measures to lower the
process temperature