4 Transmitter
67
4.4.6.3.
Editing Main Menu Channel 3 (if this option is available)
See channel 1.
Pos: 72 /TD/Produkt verwenden/MUF 63xx/Hauptmenü Alarm bearbeiten 638x @ 3\mod_1234518541385_79.doc @ 25341 @ 455
4.4.6.4.
Editing Main Menu Alarm
With the alarm, the relays, available as options, are programmed.
In addition, the alarm statuses are shown on the display (top right)
(even without relays).
You can choose whether the alarm is to be used to monitor limit
values or as a collective alarm. If an alarm is to be used to monitor
limit values, you can choose between monitoring the minimum or
maximum value and set a limit value and hysteresis for each alarm.
In addition, every alarm can be linked to a clearly visible visual
alarm (display background lighting flashes).
An alarm delay between 0 and 240 seconds can still be assigned to
every alarm used for limit value monitoring so that both the
corresponding relay effect and the visual alarm are delayed. If the
alarm status goes out within the set alarm delay time, neither the
visual alarm nor a relay connection is triggered. .
With an alarm status present, the visual alarm and all relay outputs
can be reset by means of acknowledgement. The triggering of a
new alarm cannot be enabled until after the alarm status goes out.
1.
In the Measuring Mode press
SET
, select
Main Menu Alarm
with
X
or
S
and confirm selection with
SET
.
-
Four alarms can be parameterized.
2.
Select
Alarm x
with
X
or
S
and confirm selection with
SET
.
Using alarm to monitor limit values
NO contact
Monitoring minimum
Monitoring maximum
Hysteresis
Hysteresis
Limit value
Limit value