4 Transmitter
78
to the right using
X
and increasing the value of digit by 1 using
S
. Confirm with
SET
or abort entry with
ESC
.
5.
Use
X
or
S
to select
Adj. Point 2
.
6.
Press
SET
. Read off multimeter display (e.g. 12,001 mA) and
enter this value in the user menu. Do this by scrolling one digit
to the right using
X
and increasing the value of digit by 1 using
S
. Confirm with
SET
or abort entry with
ESC
.
7.
Use
X
or
S
to select
Adj. Point 3
.
8.
Press
SET
. Read off multimeter display (e.g. 18,401 mA) and
enter this value in the user menu. Do this by scrolling one digit
to the right using
S
and increasing the value of digit by 1 using
X
. Confirm with
SET
or abort entry with
ESC
.
9.
With
X
or
S
continue to
Analog Adj. Ch. 2
or
3
(repeat steps 3
to 8).
10.
Return to main menu
Adjust
with
ESC
.
11.
Continue to main menu
Reset
with
X
or
S
or return to
Measuring Mode with
ESC
.
Performing the pressure adjustment
Please refer to n-point adjustment (pressure), page 56
1.
In the Measuring Mode press
SET
, select main menu
Adjust
with
X
or
S
and confirm selection with
SET
.
The pressure adjustment can be performed at 3 to 6
adjustment points.
2.
Select
Adjust Pressure
with
X
or
S
and confirm with
SET
.
3.
Use
X
or
S
to select
Adj. Point 1
.
4.
Press
SET
. Read off pressure sensor display (e.g. 30.1 Pa) and
enter this value in the user menu. Do this by scrolling one digit
to the right using
X
and increasing the value of digit by 1 using
S
. Confirm with
SET
or abort entry with
ESC
.
5.
Repeat steps 1 to 4 for adjustment points 2 to 6.
6.
Return to main menu
Adjust
with
ESC
.
7.
Continue to main menu
Reset
with
X
or
S
or return to
Measuring Mode with
ESC
.