4 Transmitter
84
Message Display
Cause
Remedying of fault
01116
High adjustment
temperature
The ambient
temperature is too high
during the pressure
adjustment
Take necessary
measures to lower
ambient temperature,
e.g. through venting.
03000
Heat function
defective
For testo 6614 probe
only: Heat function
defective
Contact Testo Service
Pos: 84 /TD/Produkt verwenden/MUF 63xx/Status-Warnmeldungen/Behandlung von Alarmmeldungen 638x @ 3\mod_1234769573888_79.doc @ 25573 @ 3
4.5.4.
Handling alarm messages
Shown on the
display
7
Can be used for
collective alarm
8
Additional
message end
9
New limit value
x
Scaling changed
x
Pressure too high
x
x
Delta p adjustment
x
Alarm 1
x
Alarm 2
x
Alarm 3
x
Alarm 4
x
Transmitter reset
x
Reset Min/Max
Transmitter refresh
x
User Setting Change
Analog adjustment
x
7
If multiple messages/alarms are activated at the same time, only the last
message/alarm is shown. If this is cancelled, the other messages that are still
active are no longer displayed.
8
The message can be assigned the collective alarm function, which means that
the collective alarm is activated as soon as at least one of the messages
assigned to it is activated. The collective alarm can be assigned to each of the 4
optional relays. The collective alarm is then always the same, as it can only be
defined once.
9
The message is shown upon the occurrence of the event causing the message
as well as when closing. Two entries appear in the history in the P2A software:
Message text_start and Message text_end.