5 Parameterizing, adjusting and analyzing software (P2A software)
25
Field Explanation
Unit/
Analog output
All analog outputs are parameterized in this
mask.
Unit/analog output
(graphic)
Unit: 0 to 1 V/5 V/10 V or 4 to 20 mA.
Vertical: Current version of the analog output
(cannot be changed).
Horizontal: Min./max. scale end points of
selected unit.
The curve changes in accordance with the
entered value of scale minimum and
maximum.
Scale minimum/
maximum
The endpoints of the scaling can be selected
up to the stored scale minimum and
maximum. In the process, scaling can take
place beyond the measuring range in order
to adjust the analog output to the customer
system, see Scaling page 12.
Unit
Selection of the physical unit.
When changing the unit, standard values are
set for scale minimum and maximum.
Signal delay
(graphic)
Curve changes according to the set signal
delay.