4 Transmitter
8
Accuracy
1
The specifications are only valid if the positive pressure is
applied at the positive pressure connection.
• 1.2 % of measuring range, additional ±0.3 Pa intrinsic error
2
• T
K slope drift
= 0.05 % of measuring range per degree Kelvin of
deviation from nominal temperature 22 °C
• T
K zeroing drift
= 0 % (by means of automatic zeroing)
Measuring range, resolution and overload
Measuring range
depending on
version ordered
Resolution Overload
0 to 100 Pa
0.1 Pa
20,000 Pa
0 to 10 hPa
0.01 hPa
200 hPa
0 to 20 hPa
0.01 hPa
200 hPa
0 to 50 hPa
0.01 hPa
750 hPa
0 to 100 hPa
0.1 hPa
750 hPa
0 to 500 hPa
0.1 hPa
2500 hPa
0 to 1000 hPa
1 hPa
2500 hPa
0 to 2000 hPa
1 hPa
2500 hPa
-100 to 100 Pa
0.1 Pa
20,000 Pa
-10 to 10 hPa
0.01 hPa
200 hPa
-20 to 20 hPa
0.01 hPa
200 hPa
1
At 25 °C. A stabilization period of approx. 30 min must be taken into account
upon commissioning the instrument.
2
Measuring uncertainty in accordance with GUM: ±1.2 % of measuring range
final value ±0.3 Pa.
GUM (Guide to the Expression of Uncertainty in Measurement): ISO guideline for determining the
measuring uncertainty in order to render global measurement results comparable.
The following uncertainties are used during the inquiry:
• Hysteresis
• Linearity
• Reproducibility
• Long-term
stability
•
Adjustment area/factory calibration
• Test
location