8 Tips and assistance
80
Description of
problem
Possible causes / solution
The cover of the
particulate matter
sensor cannot be
opened to clean the
sensor module.
The particulate matter sensor has already
established pressure of 400 mbar.
> If the sensor module needs to be cleaned, switch
off the sensor pump:
[ ]
→
Measurement options
→
[OK]
→
Particle
→
[OK]
→
Options
→
Particle sensor
pump: on/off
When creating a new
customer, the
Solid
Fuel
menu is not
displayed.
The testo 330 is not connected to the testo 380.
> Insert the testo 330 into the testo 380 holder.
The
Solid Fuel
menu is selectable and the
master data is transferred.
If another fuel is selected for a
measurement, it is automatically transferred
to the master data.
On the printout the
furnace setting
Combination
Boiler
/
Air Heater
is
not correct.
Master data is not created correctly.
> In
the
Address/Location
menu customise the
furnace setting
Combination Boiler
/
Air Heater
.
If we have not been able to answer your question, please contact
your dealer or Testo Customer Service. For contact details, see the
back of this document or visit the website www.testo.com/service-
contact.
8.2.
Accessories and spare parts
Description
Item no.
Cleaning kit
0554 0237
Sensor module
0394 0001
Nozzle 0394
0002
Fixing chain for particulate matter probe
0554 9356
Cleaning brush for probe shaft
0554 0228
Replacement dirt filter for condensate trap
0554 3381
Cover cap for condensate trap
0192 0099
Cover cap for probe shaft
0192 2455