7 Maintaining the product
62
9.
Close the rotation diluter cover (turn clockwise).
10.
Insert the particulate matter probe into the holder and secure
with the rubber ring.
7.8.
Cleaning the probe shaft of the particulate
matter probe
In the event of any visible contamination, the probe shaft must be
cleaned.
CAUTION
Hot probe shaft!
Risk of burns!
> Allow the probe shaft to cool down before you touch or pack it!
1.
Take the particulate matter probe out of the holder in the
transport case.
2.
Open the rotation diluter cover (turn anti-clockwise).
Содержание 0632 3800