6 Using the product
34
4.
Once the tightness test has been successfully conducted (both
traffic lights on green), remove the cover caps from the fresh air
inlet and probe shaft.
5. Finish the tightness test:
[Close]
.
6. Put the cover caps back in the plastic bag.
6.2.2.
Particle: test types – compliance measurement,
adjustment help and approval test
To prevent incorrect measurements, the position of the
testo 380 and the particulate matter probe must not be
changed during a measurement.
For flue gas loss measurement, Testo recommends using
a combustion air temperature probe with connecting cable
(0600 9787). Self-heating of the measurement system
during operation may influence the combustion air
temperature measurement with a mini ambient air probe.
Call up function:
>
[ ]
→
Measurement options
→
[OK]
→
Particle
→
[OK]
.
The testo 380 has 3 test types:
• the
Compliance measurement
test is for official
measurements in accordance with the 1st BImSchV
• the
Adjustment help
test is for the adjustment of combustion
processes
•
Approval test
is for individual room furnaces
> The operating status of the combustion system to be measured
must comply with the national directives.
Options
>
[Options]
→
Tightness test
: the tightness of the fresh air and
gas path is tested.
>
[Options]
→
Operation check
: an operation check of the
measurement system can be carried out, see
Operation check,
page
74
.
>
[Options]
→
Instrument information particle probe
:
information on the particulate matter probe is displayed.
>
[Options]
→
Instrument information particle analyzer
: testo
380 information is displayed.
>
[Options]
→
Edit cone data
: select and edit nozzle calibration
data.