6 Using the product
46
6.4.
Concluding measurement
1. Switch off the testo 330: Press
[
]
.
2.
Switch off the testo 380:
[
]
.
3. Remove crude gas tube from the particulate matter probe
connection.
ATTENTION
Backflow of condensate from the crude gas tube!
Clogging of the rotation discs and damage to the testo 380.
> Before removing the particulate matter probe from the flue gas
duct, remove the crude gas tube from the particulate matter
probe connection and dispose of the condensate in a suitable
container.
CAUTION
Hot probe shaft!
Risk of burns!
> Allow the probe shaft to cool down before you touch or pack it!
4. Remove the particulate matter probe connections
- to the probe socket
- to the particulate matter sensor with preheating section
- to the cotton filter.