32
9.6.9 Colour of the characters in the display
►
Select [coL�F] for volumetric flow or [coL�T] for temperature or [col�P] for
pressure and set the colour of the characters for the process value in
the standard display:
-bk/wh = permanently black/white
-yellow = permanently yellow
-green = permanently green
-red = permanently red
-r-cF = display colour red between the limits cFL���cFH, outside
colour change to green
-G-cF = display colour green between the limits cFL���cFH, outside
colour change to red
►
Select [cFH�x] and [cFL�x] and set the limits for the colour window:
-cFH�F = upper limit for volumetric flow
-cFL�F = lower limit value for volumetric flow
-cFH�T = upper limit value for temperature
-cFL�T = lower limit value for temperature
-cFH�P = upper limit for pressure
-cFL�P = lower limit for pressure
►
Select [coL�V] and set the colour of the characters for totaliser:
-bk/wh = permanently black/white
-yellow = permanently yellow
-green = permanently green
-red = permanently red
Menu COLR:
[coL�x]
[cFH�x]
[cFL�x]
[coL�V]
9.6.10 Error behaviour of the outputs OUT1 or OUT2
►
Select [FOUx] and set error behaviour:
-On
=
the analogue signal goes to the upper error value (→ 3.2)
-OFF =
the analogue value goes to the lower error value (→ 3.2)
-OU = the analogue signal goes to 21�5 mA in case of a pressure
measurement error and to 3�5 mA in case of a flow or tem-
perature measurement error
(→ 3.2)
Menu OUTx:
[FOUx]