AMETEK CTS
NSG 3040A, 3060A series
Operating Manual
V 1.04
100 / 157
The ANSI coupling is only implemented in NSG 3060A models with built in surge module
ANSI A coupling after n pulses
The coupling mode with 12
Ω und 9 µF will automatically be
changed after the preselected number of pulses has been
released. All possible coupling modes will be selected. The
same parameters as under Quick Start can be selected.
ANSI B coupling after n pulses
The coupling mode with 2
Ω und 18 µF will automatically be
changed after the preselected number of pulses has been
released. All possible coupling modes will be selected. The
same parameters as under Quick Start can be selected.
For coupling ANSI B
coupling L-N
is not
supported
with
NSG system.
Surge Generator and coupling impedances
Line to Line
Line to PE
(IEC)
Line to GND
(ANSI
)
IEC
2
Ω; 18 µF
12
Ω; 9 µF
ANSI A
12
Ω; 9 µF
12
Ω; 9 µF
ANSI B
2
Ω; 18 µF
2
Ω; 18 µF