11.6. Burst Test Setup
- The test generator and the coupling network should be connected to the reference ground plane (acc. to high
frequency requirements).
- The equipment under test must be isolated from the reference ground plane. The distance should be 10 cm.
Being part of the EUT, these requirements are also recommended for all connected cables. The EUT should
only be grounded if this is recommended by the installation guideline. For safety reasons, the test without any
ground connection should be conducted as well (at 100 MHz 1 m ground cable length has an impedance of
about 600
Ω
)
- Whenever possible the test setup and the cabling should always be the same; e.g. for testing power lines it
would be possible to fix the cables on the test table for all tests in the same way.
- Lines under test and all other lines should be decoupled strictly.
Coupling from NSG 3040A / 3060A generator
Coupling via 3-ph CDN to a 3-ph EUT