AMETEK CTS
NSG 3040A, 3060A series
Operating Manual
V 1.04
40 / 157
5.
Design and Functions
5.1.
Front view NSG 3040A and NSG 3060A
1
ACTIVE indication
2
Touch display
3
Rotary wheel (Inc. / Dec /Enter)
4
"Test On"
5
HV pulse Burst output 50
6
Ground reference (calibration)
7
EUT test supply
8
CRO Trigger output
5V
9
Start / Pause
10
Stop
11
Back
1
ACTIVE
Indicates that this unit is active
2
Touch screen and user interface
With the 7” capacitive touch screen display the user controls the equipment. The display indicates the status of
the generator.
3
Rotary wheel (Inc. / Dec / Enter)
Rotate for increments or decrements a selected parameter. Push change the digit and hold with a numeric
value or selects from a list of parameters. Press to change adjustment range. Selected range will be shown
briefly next to selected parameter. Push and hold to “zero” to nearest range.
4
Test On
After pressing the "Test On" key, the following actions are activated:
- Activate the high voltage power unit; HV is ready to start.
- Switches ON the supply to the EUT test supply (only with built in Power Fail module).
5
HV pulse burst output 50
External coupling devices such as the capacitive coupling clamp and the 3-phase coupling networks are
connected to the coaxial 50
Ω output. Also, the calibration of the generator is handled at this output.
Note: The burst signal is present on this output at every coupling mode.
6
Ground reference
The ground reference (4mm) is used for EFT/Burst calibration adapters.
7
EUT test supply
For single-phase EUT the coupling/decoupling network is part of the generator. The EUT is powered via the
safety banana plugs at the front panel of the simulator. Rated power max. 300V AC/DC 16 A
Attention
For verification or impulse measuring, you must take care, to use measuring instruments,
which are able to handle the high voltages up to 6.6 kV and surge currents up to 3300 A.