AMETEK CTS
NSG 3040A, 3060A series
Operating Manual
V 1.04
138 / 157
16.2. Telecom Surge pulse settings
16.2.1.
Configuration
In the configuration menu the user set the following parameters:
-
Signal
Trigger out signal
at the BNC plug on generator front side
-
Start trigger
for the impulse or event when Start button is blinking
- Behavior of the
EUT Monitor 1
on BNC plug on rear side
16.2.1.1.
Signal Trigger
The trigger out signal is a programmable 5V edge signal located at the front side BNC Trigger plug. This
signal is for trigger other devices like oscilloscopes or other devices.
If the start trigger is not set to Automatic, the trigger indicates with “
Waiting for trigger
”
the recommended
trigger signal. The waiting time for a trigger signal is endless if no trigger timeout is enabled.
The generator offers the following status:
-
Automatic
- At
Start
(starts after the trigger)
- Each
pulse
(same as trigger out)
- Each
step
(same as trigger out)
Iterations
For Iteration tests the parameters changes after each step as shown in the figure below. Depends of the
order and setting the following parameters can change:
- Polarity
- Voltage
Example with voltage iteration that illustrate trigger out for cycle and step