Getting Started
30
MIRA3 FEG-SEM
Note: Contact between the specimen and any part of the chamber is indicated by a special
electric circuit. This circuit is based on the measuring the electric current. It works only if the
specimen is conductive.
6.
Close the chamber door by pushing it towards the chamber.
7.
Start the pumping procedure by clicking on the PUMP button in the Vacuum panel,
where the actual pressure and pumping progress are also displayed. Ensure that the
chamber door is tightly closed.
8.3
Images at Low Magnification
There are four factory presets for the accelerating voltage (2, 5, 10, 20) kV, one for each HV
index. The user does not need to perform any further adjustments by switching between
them and using magnification up to 4000x.
1.
In the SEM Detectors & Mixer panel select the appropriate detector from the list box
(Figure 8). Using the SE or BSE detector is recommended. When the BSE detector is
used, ensure that the detector is not retracted! See chapter 6 for detailed
information.
Figure 8
Note: Notice the difference between SE and BSE images below (Figure 9).
Figure 9
SE Detector
BSE Detector