Description of the Microscope
16
MIRA3 FEG-SEM
panel. Press Next>> button. Minimize the image movement by changing the IML
Centering (select in the Pad panel) using the trackball and the F11 and F12 keys.
Press Next>> button. Adjust the OBJ Centering so that the image is the same
as in RESOLUTION mode (use the trackball and the F11 and F12 keys).
Press the
Finish button.
6.
Switch to DEPTH mode. Use again the procedure with Manual Column Centering and
OBJ Centering (steps 3. and 4.).
7.
Switch to WIDE FIELD mode and set the same magnification for RESOLUTION and
WIDE FIELD modes. Use the Adjustment >>> Manual Column Centering function
in the Electron Beam panel. Press Next>> button. Adjust the OBJ Centering so that
the image is the same as in RESOLUTION mode (use the trackball and the F11 and
F12 keys). Press the Finish button.
MIRA3 with InBeam detector:
It is recommended to work with enabled InBeam mode all the time. However, there is the
possibility to use purely magnetic objective lens. Proceed with following:
1.
Uncheck the item InBeam Mode in the main SEM menu.
2.
Select the SE detector in the SEM Detectors & Mixer panel.
3.
Switch to RESOLUTION mode and center the mode as usual. Ensure that the image is
correctly stigmated; especially at low accelerating voltages (it is applicable also for
other modes).
Note: If the InBeam mode is switched on/off, it is recommended to carry out the centering
of the electron column again.
It might also be necessary to have FIELD or DEPTH modes precisely centered if working
at higher magnifications. If needed proceed as follows:
4.
Switch to DEPTH mode and centre this mode as usual.
5.
Switch to FIELD mode and only set the same field of view by changing the OBJ
Centering.
6.
Switch to WIDE FIELD mode and center this mode as usual.
7.
Repeat steps from 3. to 6. for each of four HV indexes.
Note: If you want to align field of view for DEPTH mode exactly with RESOLUTION mode (for
magnifications > ~10kx), set the same magnification in RESOLUTION and DEPTH modes and
then set the same field of view by means of changing the IML Centering parameter.
4.1.2.4
Electron Column Precentering
The column is factory precentered. It is also precentered after every filament exchange
by the service technician. Further precentering can be performed only by an Expert level
user. This can be an advantage at lower accelerating voltages (HV < 5 kV). On the other
hand, incorrect precentering can cause a situation where the objective and stigmators cannot
be corrected at all.
WARNING: This procedure has a strong influence on the other centering parameters and thus
is recommended only for experienced user.