Description of the Microscope
14
MIRA3 FEG-SEM
4.1.2
Electron Column Centering
To reach high quality images it is necessary to center the electron optics of the column. The
centering is electronic and fully controlled by the PC.
Each microscope is factory centered and this setup is stored in the Default configuration. By
using the Load Default Configuration function from the menu Options -> Configurations
►
the user will save time in the case he gets lost, loses his configuration completely etc. This
default configuration file is also used when a new user is created as his initial configuration.
There are factory presets for four accelerating voltages, one for each HV index (see the table
on page 16). Switching among them the MiraTC software remembers all settings so the user
does not need to make any further adjustments. However, changing the voltage within
a single HV index requires performing the centering of the microscope:
1.
centering of the gun (see chapter 4.1.2.1);
2.
centering of the column, either automatic (see chapter 4.1.2.2) or manual (see
chapter 4.1.2.3) for work at high magnifications.
Note: Every user can save his own configurations in menu Options -> Configurations
►
.
Save
as. Using of various configurations can save user’s time when switching between various
voltages within the same HV index.
Recommended centering conditions
The centering is always done by means of a special centering specimen, which is included
in the microscope accessories and is marked as ADJ. The surface of this specimen has a fine
structure, making the centering easier. If this specimen is not available, it is possible to use
a clean common specimen stub. Once you start centering, the image must be focused
perfectly.
Mode
Working Distance (WD)
Beam Intensity (BI)
Magnification (Mag)
RESOLUTION Current WD on what you
want to acquire images
Current BI on what you
want to acquire images
(2000–5000)x
DEPTH
Current WD
Current BI
2000x
FIELD
Current WD
Current BI
500x
WIDE FIELD
Current WD
Current BI
max. Mag
4.1.2.1
Centering of the Electron Gun
The centering is done by automatic procedure. It is always necessary to perform this auto
centering in each HV index which is used. Use HV according to the table on page 17.
1.
Insert the centering specimen.
2.
Focus the image in RESOLUTION mode and then set up minimum magnification (use
WD approximately 5 mm or any preferable WD).