4) Set Cycle Time
Cycle time is adjustable from 5 to 30 seconds.
Press [S] button cycle to enter this menu, and then press the [
▲] and [▼]
button to adjusting the setting.
4.7.4. MICROCURRENT Setting
Press the [S] button cycle to enter the setting state. The settings can be adjusted
according to the following steps:
1) Set the Therapeutic Program
There are 3 programs in MIC therapeutic program available
—Constant, Pulse Width Modulation, and Pulse Rate Modulation.
The therapeutic program can be selected by pressing the [
▲] and [▼] button.
The mode you selected will show up on the top of liquid crystal display.
2) Set Timer
Press [S] button cycle to enter this setting. The treatment time is adjustable
from 1 to 60 minutes or Continuous. Press [
▲] or [▼] button control to adjust
setting. You can set the timer to
“Continuous” mode by pressing the [▲] control
when it shows 60 minutes. Its output will be shut off when time is up.
3) Set Pulse Width
Pulse Width is adjustable from 2ms to 200ms
. Press “S” button cycle to enter
this menu, then press [
▲] or [▼] button to adjust the setting.
4) Set Pulse Rate
Pulse rate is adjustable from 1 Hz to 150Hz. Press [S] button cycle to enter this
menu, and then press [
▲] or [▼] button to adjust the setting.