20
Caution:
1)
Battery may be fatal if swallowed. Therefore, keep the battery and the product
out of the range of children. If a battery was swallowed, consult a physician
immediately.
2) If a battery has leaked, avoid contact with skin, eyes and mucus membranes,
Rinse the affected spots with lots of clear water immediately and contact a
physician right away.
3) Battery may not be charged, dismantled, thrown into fire or short-circuited.
4) Protect battery from excess heat; Take the batteries out of the product if they
are spent or in case you no longer use the article. This prevents damage caused
by leaking battery.
5) Always replace with the same type battery.
4.2 Connect electrodes to lead wires
Insert the lead wire connector into electrode connector (standard 2mm female
connection). Make sure that no bare metal of the pins is left exposed.
Caution:
Always use electrodes that meet local safety standard:
CE mark: with IEC/EN6060I-1, ISO10
993-1/-5/-10
and
lEC/
EN60601-1-2
FDA: 510(K) certified.
4.3 Connect lead wires to device
1) Ensure that the device is completely turned OFF.
2) The wires provided with the system insert into the jack sockets located on top of the
device.
3) Holding the insulated portion of the connector, push the plug end of the wire into one
of the jacks (see drawing). One or two sets of wires may be used.