Preparing the Pump for Operation
Using the Data System to Start and Stop the Solvent Flow
38
Accela Pump Hardware Manual
Thermo Scientific
Figure 29.
Using a syringe to draw solvent through a solvent reservoir line
3. Connect a 10 mL syringe to the Luer-Lok fitting, and pull solvent through the line.
4. Remove the Luer-Lok fitting from the port.
5. Reconnect the solvent line to the degasser outlet port.
To remove air from the LDA and from the solvent lines that connect the degassing unit to the
proportioning valves, start the solvent flow by using the data system controls (see
“Using the
Data System to Start and Stop the Solvent Flow”
on
page 38
).
Using the Data System to Start and Stop the Solvent Flow
You can start and stop the solvent flow by using the data system start and stop controls:
•
Using Xcalibur to Start and Stop the Solvent Flow
•
Using ChromQuest to Start and Stop the Solvent Flow