6
Maintenance
Removing the LDA from the Pump
Thermo Scientific
Accela Pump Hardware Manual
61
Figure 39.
View of the high-pressure tubing connections
5. Unscrew the low-pressure solvent lines connected to the solvent proportioning valves.
6. Using a flat blade screwdriver, loosen the small, nylon screws that secure the actuating
coils to the proportioning valves, and then pull the actuating coils off the ends of the
proportioning valves (see
Figure 40
).
Figure 40.
View of the actuating coils with their nylon screws
A
B
C
D
PU RGE
PULSE
DAMPENER
FLUSH
Crossover
tubing
Connection to
outlet module
Captive screws (3x)
CAUTION
Take care not to pinch the wires that connect the actuating coils to the
pump. Pinching a connecting wire can irreversibly damage the thin electrical wire
housed within the white insulated coating. The damage might not be apparent until
you attempt to run the pump.
Nylon screw
White, insulated, electrical wire