T3PSX3200P Series User Manual
154
:TRIGger:OUT:CONDition
{D0|D1|D2|D3|D4},{OUTOFF|OUTON|>V|<V|=V|>C|<C|
=C|>P|<P|=P|AUTO},<value>
Description
Sets the trigger condition from the trigger output
of the specified data line.
Parameter
OUTOFF|OUTON|>V|<V|=V|>C|<C|=C|>P
|<P|=P|AUTO
When OUTOFF, OUTON or AUTO is selected,
parameter <value> will be omitted.
Input trigger: disable output (OUTOFF) and
enable output (OUTON), both of which indicate
to disable or enable respectively the trigger when
the specified controlled channel output.
Voltage trigger: greater than voltage (>V), less
than voltage (<V), equal to voltage (=V), all of
which indicate that when the output voltage of
the specified controlled channel meets the set
trigger condition, it will trigger then.
Current trigger: greater than current (>C), less
than current (<C), equal to current (=C), all of
which indicate that when the output current of
the specified controlled channel meets the set
trigger condition, it will trigger then.
Power trigger: greater than power (>P), less than
power (<P), equal to power (=P), all of which
indicate that when the output power of the
specified controlled channel meets the set trigger
condition, it will trigger then.
Automatic trigger: instrument automatically
triggers after enabling this function.
When either one of the following modes: voltage
trigger (>V, <V, =V), current trigger (>C, <C, =C),
Содержание T3PS13206P
Страница 1: ... ...
Страница 106: ...T3PSX3200P Series User Manual 106 Register Commands OPC 187 OPC 187 ...
Страница 177: ...REMOTE CONTROL 177 Example SAV 1 Recalls the setting stored in memory 2 STATE02 ...
Страница 185: ...REMOTE CONTROL 185 Example STB Returns 81 if the status byte register is set to 0101 0001 ...
Страница 196: ...T3PSX3200P Series User Manual 196 Between chassis and DC power cord 30MΩ or above DC 500V ...
Страница 197: ... 0 0 0 1 2 3 4 5 6 7 8 8 9 3 3 3 45 6 990 0 0 9 0 0 0 9 0 0 0 0 9 7 7 88 8 1 931711 RevB ...