144
:SEQUence[1|2]:TEMPlet:INTErval <value>
Description
Sets the time interval.
Parameter
<value> Integer 1s to 300s
Time interval is the duration period in
accord with each time parameter of the
active selected templet.
Example
:SEQUence2:TEMPlet:INTErval 15
Sets the CH2 time interval as 15s. Pulse templet
does Not support this parameter
.
:SEQUence[1|2]:TEMPlet:INTErval?
Description
Queries the active time interval.
Example
:SEQUence2:TEMPlet:INTErval?
Queries the CH2 active time interval. Returns to
an integer between 1 to 300.
:SEQUence[1|2]:TEMPlet:INVErt {ON|OFF}
Description
Enable or disable invert function of the active
selected templet.
Parameter
When enabling the invert function, invert the
target waveform in instrument first, and build
timer parameter. Invert function is available for
Sine, Pulse and Ramp templets only.
Example
:SEQUence2:TEMPlet:INVErt ON
Enable the invert function of active selected
templet for CH2.