76
Teledyne API N500 CAPS NO
X
Analyzer with NumaView™ Software
089770000 Rev 05
WARNING
FAULT CONDITION
POSSIBLE CAUSES
CELL PRESS WARN
Cell pressure is too high or too
low for accurate NO, NO
2
, NO
x
readings.
(<5 in-Hg or > 35 in-Hg).
Normally 29.92 in-Hg at sea level
decreasing at 1 in-Hg per 1000 ft
of altitude (with no flow – pump
disconnected).
If Cell Pressure is < 5 in-Hg:
•
Failed Pressure Sensor/circuitry
If Cell Pressure is > 35 in-Hg:
•
Bad Pressure Sensor/circuitry
•
Pressure too high at Sample Inlet.
CONFIG INITIALIZED
Configuration and Calibration
data reset to original Factory
state or erased.
Failed Disk on Module
User erased data
DATA INITIALIZED
Data Storage in DAS was erased
before the last power up
occurred.
Failed Disk-on-Module
User cleared data.
SAMPLE PRESS WARN
Sample pressure is too high or
too low for accurate NO, NO
2
,
NO
x
readings.
(<15 in-Hg or > 35 in-Hg).
Normally 29.92 in-Hg at sea level
decreasing at 1 in-Hg per 1000 ft
of altitude (with no flow – pump
disconnected).
If Sample Pressure is < 15 in-Hg:
•
Blocked Particulate Filter
•
Blocked Sample Inlet/Gas Line
•
Failed Pressure Sensor/circuitry
If Sample Pressure is > 35 in-Hg:
•
Bad Pressure Sensor/circuitry
•
Pressure too high at Sample Inlet.
SYSTEM RESET
The computer has rebooted.
This message occurs at power on.
If it is confirmed that power has not been
interrupted:
5 VDC power
Fatal Error caused software to restart
Loose connector/wiring
1
Clears the next time successful zero calibration is performed.
2
Clears the next time successful span calibration is performed.
FAULT DIAGNOSIS WITH DASHBOARD FUNCTIONS
In addition to being useful as predictive diagnostic tools, the functions viewable in the
Dashboard can be used to isolate and identify many operational problems.
The acceptable ranges for these functions are listed in the “Nominal Range” column of
the analyzer
Final Test and Validation Data Sheet
shipped with the instrument. Values
outside these acceptable ranges Indicates a failure of one or more of the analyzer’s
subsystems. Functions whose values are still within acceptable ranges but have
significantly changed from the measurement recorded on the factory data sheet may also
Indicates a failure.
Make note of these values for reference in troubleshooting.
USING THE DIAGNOSTIC SIGNAL I/O FUNCTIONS
The signal I/O functions in the Utilities>Diagnostics menu allows access to the digital
and analog I/O in the analyzer. Some of the digital signals can be controlled through the
Setup menu. These signals are useful for troubleshooting in three ways: