Teledyne API N500 CAPS NO
X
Analyzer with NumaView™ Software
089770000 Rev 05
MODE
DESCRIPTION
Modem Init String
Input an initialization string to enable the modem to communicate.
Parity
Select odd, or even, or no parity (typically set in conjunction with Data Bits and
Stop Bits).
Protocol
Select between the communications protocols: MODBUS RTU or MODBUS ASCII
Quiet Mode
Enable/disable Quiet mode, which suppresses any feedback from the analyzer
(such as warning messages) to the remote device and is typically used when the
port is communicating with a computer program where such intermittent
messages might cause communication problems.
Such feedback is still available, but a command must be issued to receive them.
Security
Enable/disable the requirement for a password for this serial port to respond. The
only command that is active is the request-for-help command (? CR).
Stop bits
Select either 0 or 1 stop bit (typically set in conjunction with Parity and Data bits).
TCP
P
ORT
1
TCP Port1 allows choosing whether or not to display the command prompt, editing the
Port 1 number for defining the terminal control port by which terminal emulation
software addresses the instrument, such as Internet or NumaView™ Remote software,
and enabling or disabling security on this port.
TCP
P
ORT
2
TCP Port2 is configured with the port number for MODBUS.
N
ETWORK
S
ETTINGS
The Setup>Comm>Network Settings menu is for Ethernet configuration. The address
settings default to automatic configuration by Dynamic Host Configuration Protocol
(DHCP). Most users will want to configure the instrument with a static IP address: click
the Static radio button to manually assign a static IP address (consult your network
administrator and see Table 2-11 for information).