16
Teledyne API N500 CAPS NO
X
Analyzer with NumaView™ Software
089770000 Rev 05
REAR PANEL
Figure 2-2 shows the layout of the rear panel.
Figure 2-2. Rear Panel Layout
Table 2-2. Rear Panel Description
COMPONENT
FUNCTION
Model/specs label
Identifies the analyzer model number and provides power specs
SAMPLE
Connect a gas line from the source of sample gas here.
Calibration gases can also enter here on units without zero/span/shutoff valve options
installed.
EXHAUST
Connect an exhaust gas line of not more than 10 meters long here that leads outside the
shelter or immediate area surrounding the instrument. The line must be ¼” tubing or
greater.
SPAN
On units with zero/span valve option installed, connect a gas line to the source of
calibrated span gas here.
ZERO AIR
On units with zero/span valve option installed, attach a gas line to the source of zero air
here. If a permeation oven, also known as internal zero/span valve (IZS), option is
installed attach the zero air scrubber here.
cooling fan
Pulls ambient air into chassis through side vents and exhausts through rear.
Power
cable
connector
Connector for three-prong cord to apply AC power to the analyzer.
CAUTION! The cord’s power specifications (specs) MUST comply with the power
specs on the analyzer’s rear panel label
Power
On/Off
Switch
Rocker switch to power unit on or off
CAUTION!
Prior to powering OFF, use front panel button for “soft” power-down of internal
components.
ANALOG OUT Option
For voltage or current loop outputs to a strip chart recorder and/or a data logger.
DIGITAL I/O Option
For remotely activating the zero and span calibration modes.
ETHERNET
Connector for network or Internet remote communication, using Ethernet cable
COM
Serial communications port for RS-232 or RS-485