66
Teledyne API N500 CAPS NO
X
Analyzer with NumaView™ Software
089770000 Rev 05
Table 5-2. Predictive Uses for Dashboard Functions
FUNCTION EXPECTED
ACTUAL
INTERPRETATION & ACTION
Cell Press
(pressure)
Constant within
atmospheric
changes
Fluctuating
Developing leak in pneumatic system. Check for leaks.
Slowly increasing
Flow path is clogging up. Replace orifice filters.
Slowly decreasing
Developing leak in pneumatic system to vacuum (developing valve
failure). Check for leaks.
AREF
Constant within
±100 Mm-1 of
check-out value
Significantly increasing
Developing AREF valve failure. Replace valve.
Developing leak in pneumatic system. Check for leaks.
Debris on mirrors. Contact Technical Support.
Gas Conc
(concentration)
Constant for
known gas
concentration
Decreasing over time
Developing leak in pneumatic system. Check for leaks.
OPERATIONAL HEALTH CHECKS
Navigate to the Utilities>USB Utilities>Report menu (Figure 5-1) to download a report
on the basic operations of the instrument. To download the report for your own viewing
on a computer or to send to others, insert a flash drive into a front panel USB port and
press the Download button, which is enabled when the instrument detects the flash drive.
Figure 5-1: Report Generation Page
The report can also be set to generate periodically and sent to a Web services “cloud”
where it is available for viewing by Teledyne API technical support personnel. Set this
function with two Vars:
Setup>Vars>Upload Report to Cloud: set to True.
Setup>Vars>Report Upload>Interval: edit the number of hours between report uploads.
Содержание N500