68
Teledyne API N500 CAPS NO
X
Analyzer with NumaView™ Software
089770000 Rev 05
4. In the Update Firmware field, press the Check button for the instrument to
determine whether the firmware on the flash drive is more recent than what is
currently installed. Once it’s been determined that the firmware is new, the
Install button will be enabled; if the firmware version on the flash drive is the
same as or older than the current firmware of the instrument, the Install
button will not be enabled.
5. Press the Install button and note the messages in the Status field at the
bottom of the page. Use the Cancel button if necessary.
6. When complete, as indicated in the Status field, press the Done button,
which replaces the Cancel button, and remove the flash drive.
7. Power off and restart the instrument to complete the new firmware
installation.