5.5 Electrical Troubleshooting
If your instrument stops working and the touch panel display
is off, check the line cord connection.
If the line cord is connected properly, check the fuses in the
power input module located on the system’s rear panel. These
are time delay fuses (T-type) and rated according to the mains
input power voltage:
• 100 to 129 VAC mains input: 2.0 A fuse
• 215 to 264 VAC mains input: 1.0 A fuse.
CAUTION
Always replace fuses with the same type and rating.