Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
72
2.18
Memory Group Commands
Command
*RCL (No Query Form)
Description
This command restores the state of the instrument from a copy of
the settings stored in the setup memory.
The settings are stored using the *SAV command. If the specified
setup memory is deleted, this command causes an error.
Group
Memory
Syntax
*RCL {0|1|2|3|4}
Related Commands
*SAV
Arguments
0, 1, 2, 3, or 4
Specifies the location of setup memory
Returns
None
Example
*RCL 3
Restores the instrument from a copy of the settings stored in memory
location 3
Table 85: *RCL
Command
*SAV (No Query Form)
Description
This command stores the current AWG configuration to a specified
setup memory location.
The current AWG configuration is automatically saved every time
the user changes a parameter. This configuration, here named
"current configuration" is automatically loaded at the power on.
Note: If a specified numbered setup memory is locked, this
command causes an error.
Group
Memory
Syntax
*SAV {0|1|2|3|4}
Related Commands
*RCL
Arguments
0, 1, 2, 3, or 4
Specifies the location of setup memory
Returns
None
Example
*SAV 2
Saves the current instrument state in the memory location 2
Table 86: *SAV