Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
44
This strategy manages the length of the sequencer entries in
relationship with the length of the channel waveforms defined for
each entry
Group
Control
Syntax
AWGControl:LENGth:MODE {ADAPTLonger|ADAPTShorter|DEFault}
AWGControl:LENGth:MODE?
Related Commands
Arguments
• AD
APTLonger: the length of an entry of the sequencer by default
will be equal to the length of the longer channel waveform, among
all analog channels, assigned to the entry.
• ADAPTShorter: the length of an entry of the sequencer by default
will be equal to the length of the shorter channel waveform, among
all analog channels, assigned to the entry.
• DEFault:the length of an entry of the sequencer by default will be
equal to the value specified in the Sequencer Item Default Length
[N] parameter
Returns
ADAPTLonger|ADAPTShorter|DEFault
Example
AWGControl:LENGth:MODE ADAPTLonger sets the entry length
strategy to “Adapt to Longer Analog Waveform”
AWGControl:LENGth:MODE?
Might return DEFault
Table 36: AWGControl:LENGth:MODE
Command
AWGControl:RESET[:IMMediate] (No Query Form)
Description
This command resets sequence to its default state. The sequence
table will be cleared. Parameters such as sampling rate, skew time
will be set to its default value.
Importante Note
: the behaviour of this command is equivalent to
*RST with the exception that before sending the AWGControl:RESET
command, the AWG must be in Stopped state.
Group
Control
Syntax
AWGControl:RESET[:IMMediate]
Related Commands
Arguments
Returns
None