Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
54
Table 53: OUTPut[n][:STATe]
Command
OUTPut[n][:STATe]
Description
This command and query sets or returns the Output state of the
analog channel “n”.
Setting the output state of a channel to ON will
switch on its analog output signal.
Group
Output
Syntax
OUTPut[n][:STATe] {OFF|ON}
OUTPut[n][:STATe]?
Related Commands
None
Arguments
•
OFF: the selected output is disconnected from the DUT by
means of a relay and the analog signal is turned off.
•
ON: the selected output is turned on
The value of n indicates the channel number.
Returns
OFF|ON
Example
OUTPut1 ON
Turns on the analog output signal and closes the output relay
Command
DIGitals:LEVel[m]
Description
This command and query sets or returns the Voltage Level of the
digital output probe.
Group
Output
Syntax
DIGitals:LEVel[m] {MINimum|MAXimum|DEFault|<Volts>}
DIGitals:LEVel[m]? [{MINimum|MAXimum}]
Related Commands
None
Arguments
•
MINimum: sets the parameter to the minimum value
•
MAXimum: sets the parameter to the maximum value
•
DEFault: sets the parameter to the default value
•
< Volts > := <NRf> value
The value of m indicates the digital probe where m=1 Digital Pod A,
m=2 Digital Pod B, m=3 Digital Pod C, m=4 Digital Pod D
Returns
<NRf>
Example
DIGitals:LEVel2 2.5
Set the voltage amplitude to 2.5V for the pod B