Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
49
Syntax
AWGControl:DJStrobe <value>
Related Commands
SEQuence:ELEM[n]:PATTERN
SEQuence:ELEM[n]:PATTERNJUMPTOMode
SEQuence:ELEM[n]:PATTERNJUMPTOEntry
Arguments
<value> := <NR1> between 0 and 255
Returns
None
Example
AWGControl:DJStrobe 123
Sends 123 as pattern strobe event.
Table 46: AWGControl:DJStrobe
2.11
Calibration And Diagnostic Commands
Command
CALibration[:ALL]
Description
This command does a full calibration of the AWG. In its query form,
the command does a full calibration and returns a status indicating
the success or failure of the operation. This command is equivalent
to the *CAL? command.
Group
Control
Syntax
CALibration[:ALL]
CALibration[:ALL]?
Related Commands
*CAL?
Arguments
Returns
<calibration error code> ::= <NR1>
0 indicates no error
–
1 indicates an error
Example
CALIBRATION:ALL performs a calibration.
CALIBRATION:ALL? performs a calibration and returns results. For ex-
ample, it might return 0, indicating that the calibration completed
without any errors.
Table 47: CALibration[:ALL]
Command
DIAGnostic[:ALL]
Description
This command executes the self diagnostic procedure. The query
form of this command executes the self diagnostic procedure and
returns the results in the form of numeric of values of 0 for no errors or
-1 for one or more tests failed.