Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
62
Example
LICense:LIST?
Might return
“02f0
-4fff-b528-6ef7-f501-4515-8f38-1f54
”
Table 66: LICense:LIST?
Command
*OPT? (Query Only)
Description
This command returns the installed options and application licenses
for the AWG.
Group
License
Syntax
*OPT?
Related Commands
Arguments
Returns
<opt> ::= <option_description>: <option_value>
<option_description> ::= <string>
<option_value> ::= <string>
Example
*OPT? might return the following string:
"Memory Option: 128 MSample, Digitals Option: No Digitals,
Amplitude Option: 6 VPP"
Table 67: *OPT?
2.15
Marker Group Commands
Command
MARKer:LEVel[m]
Description
This command and query sets or returns the marker output Voltage
Level parameter in Volts.
The marker voltage is calculated on 50 ohm load; the minimum value
is 1V@50 Ohm load and the maximum value is 2.5V@50 ohm load.
Group
Marker
Syntax
MARKer:LEVel[m] {MINimum|MAXimum|DEFault|<Volts>}
MARKer:LEVel[m]? [{MINimum|MAXimum}]
Related
Commands
Arguments
•
MINimum: sets the parameter to the minimum value