BASICS
Changing the Display
The grid is 8 Vertical divisions representing 256 Vertical levels and 10
Horizontal time divisions. The value represented by each division depends
on the scale settings of the traces that appear on it.
8 Vertical
Divisions
256
Vertical
Levels
Multi-Grid Display
The grid can be divided into multiple grids each representing the full
number of Vertical levels. A variety of multi-grid modes are available.
256
Vertical
Levels
256
Vertical
Levels
By default, the oscilloscope has the
Auto
grid mode enabled. Auto adds a
grid for each new trace, up to 16 grids, until no more grids are available. Other
grid modes create a fixed number and orientation of grids; the icon shows
the result.
To modify the touch screen display, choose
Display > Display Setup
from the
menu bar and make your selections from the Display dialog.
Extended Display
If you have a second monitor connected, select
Extend Grids on 2nd
Monitor
from the Display dialog, then choose a grid mode from the Extended
Display pop-up menu. Both screens will share this grid mode. Drag-and-drop