Performance Verification: Ajustment/Verification Fixture
TLA600 Series Logic Analyzer Service Manual
4- 29
Use the following procedure to verify the internal clock frequency.
Parameter tested
Internal clock frequency
Equipment
required
Frequency counter (item 8)
Prerequisites
Warm-up time: 30 minutes for adjustment/verification fixture and test
equipment
1.
Set the jumper positions as listed in the table.
Jumper
Jumper name
Jumper setting
J13
P6434 Deskew/probe setup
3-5 connected
J15
Clock selection
INT
J16
Deskew
2-3 and 4-5 connected
J17
Setup/hold select
HOLD
J19
Clock polarity select
Removed
2.
Using the frequency counter, measure the oscillator frequency at pin J1-2.
The frequency should match what is listed on the Calibration Data report.
3.
Record this measurement on the Calibration Data Report.
Verify Internal Clock
Frequency
Содержание TLA600 Series
Страница 5: ......
Страница 13: ...Table of Contents viii TLA600 Series Logic Analyzer Service Manual...
Страница 17: ...Service Safety Summary xii TLA600 Series Logic Analyzer Service Manual...
Страница 51: ...Theory of Operation 3 4 TLA600 Series Logic Analyzer Service Manual...
Страница 89: ......
Страница 97: ...TLA600 Adjustment Procedures 5 8 TLA600 Series Logic Analyzer Service Manual...
Страница 113: ...Maintenance 6 4 TLA600 Series Logic Analyzer Service Manual...
Страница 157: ...Repackaging Instructions 6 48 TLA600 Series Logic Analyzer Service Manual...
Страница 159: ...Options 7 2 TLA600 Series Logic Analyzer Service Manual...
Страница 161: ...Electrical Parts List 8 2 TLA600 Series Logic Analyzer Service Manual...
Страница 165: ...Diagrams 9 4 TLA600 Series Logic Analyzer Service Manual...
Страница 176: ...Mechanical Parts List TLA600 Series Logic Analyzer Service Manual 10 11...
Страница 185: ...Mechanical Parts List 10 20 TLA600 Series Logic Analyzer Service Manual...
Страница 186: ......
Страница 187: ......