Specifications
1- 4
TLA600 Series Logic Analyzer Service Manual
Table 1- 3: TLA600 external signal interface (Cont.)
Characteristic
Description
Input Mode
Active (true) low, level sensitive
Input Bandwidth
1
Signal 1, 2
Signal 3, 4
50 MHz square wave minimum
10 MHz square wave minimum
Active Period
Accepts signals during valid acquisition periods via real-time gating
Maximum Input Voltage
0 to +5 V peak
System Trigger Output
TTL compatible output via rear panel mounted BNC connectors
Source Mode
Active (true) low, falling edge latched
Active Period
Outputs system trigger state during valid acquisition period, resets system trigger output to false
state between valid acquisitions
Output Levels
V
OH
V
OL
50
Ω
back terminated TTL-compatible output
≥
4 V into open circuit
≥
2 V into 50
Ω
to ground
≤
0.7 V sinking 10 ma
Output Protection
Short-circuit protected (to ground)
External Signal Output
TTL compatible outputs via rear panel mounted BNC connectors
Source Selection
Signal 1, 2, 3, 4, or 10 MHz clock
Output Modes
Level Sensitive
User definable
Active (true) low or active (true) high
Output Levels
V
OH
V
OL
50 Ohm back terminated TTL output
≥
4 V into open circuit
≥
2 V into 50
Ω
to ground
≤
0.7 V sinking 10 ma
Output Bandwidth
2
Signal 1, 2
Signal 3, 4
50 MHz square wave minimum
10 MHz square wave minimum
Active Period
Outputs signals during valid acquisition periods, resets signals to false state between valid
acquisitions
Outputs 10 MHz clock continuously
Output Protection
Short-circuit protected (to ground)
Intermodule Signal Line
Bandwidth
Signal 1, 2 (ECLTRG 0,1)
Signal 3, 4 (ECLTRG 0,1)
Minimum bandwidth up to which the signals are specified to operate correctly
50 MHz square wave minimum
10 MHz square wave minimum
1
The Input Bandwidth specification only applies to signals to the modules; it does not apply to signals applied to the
External Signal Input and sent back to the External Signal Output.
2
The Output Bandwidth specification only applies to signals from the modules; it does not apply to signals applied to the
External Signal Input and sent back to the External Signal Output.
Содержание TLA600 Series
Страница 5: ......
Страница 13: ...Table of Contents viii TLA600 Series Logic Analyzer Service Manual...
Страница 17: ...Service Safety Summary xii TLA600 Series Logic Analyzer Service Manual...
Страница 51: ...Theory of Operation 3 4 TLA600 Series Logic Analyzer Service Manual...
Страница 89: ......
Страница 97: ...TLA600 Adjustment Procedures 5 8 TLA600 Series Logic Analyzer Service Manual...
Страница 113: ...Maintenance 6 4 TLA600 Series Logic Analyzer Service Manual...
Страница 157: ...Repackaging Instructions 6 48 TLA600 Series Logic Analyzer Service Manual...
Страница 159: ...Options 7 2 TLA600 Series Logic Analyzer Service Manual...
Страница 161: ...Electrical Parts List 8 2 TLA600 Series Logic Analyzer Service Manual...
Страница 165: ...Diagrams 9 4 TLA600 Series Logic Analyzer Service Manual...
Страница 176: ...Mechanical Parts List TLA600 Series Logic Analyzer Service Manual 10 11...
Страница 185: ...Mechanical Parts List 10 20 TLA600 Series Logic Analyzer Service Manual...
Страница 186: ......
Страница 187: ......