Adjustment/Verification Fixture Adjustment
5- 20
TLA600 Series Logic Analyzer Service Manual
H
Record Length
5000
H
Horizontal Scale
Main Scale @ 500 ps/div
c.
Set up the Trigger menu as follows:
H
Source
CH2
H
Coupling
DC
H
Slope
+
H
Level
700 mV
H
Mode
Normal
d.
Set up the Measure menu as follows:
H
Select Measurement for CH1
Measure
H
Gating
Off
e.
Set up the Cursor menu as follows:
H
Function
V Bars
f.
Set up the Acquire menu as follows:
H
Acquisition Mode
Sample
H
Repetitive Signal
On
4.
Using the delay taps on Delay SETUP ADJ., add or subtract the appropriate
delay to ensure a 3.0 ns (
±
25 ps) timing difference between R67 and R127
(J4 -- SU/HOLD CLK OUT). See Note on page 5--14.
5.
Install a jumper at J19 and verify that the signal at R127 changes to a Falling
Edge.
6.
After completing the adjustment/verification fixture adjustment procedure,
do the performance verification procedure that starts on page 4--21.
Содержание TLA600 Series
Страница 5: ......
Страница 13: ...Table of Contents viii TLA600 Series Logic Analyzer Service Manual...
Страница 17: ...Service Safety Summary xii TLA600 Series Logic Analyzer Service Manual...
Страница 51: ...Theory of Operation 3 4 TLA600 Series Logic Analyzer Service Manual...
Страница 89: ......
Страница 97: ...TLA600 Adjustment Procedures 5 8 TLA600 Series Logic Analyzer Service Manual...
Страница 113: ...Maintenance 6 4 TLA600 Series Logic Analyzer Service Manual...
Страница 157: ...Repackaging Instructions 6 48 TLA600 Series Logic Analyzer Service Manual...
Страница 159: ...Options 7 2 TLA600 Series Logic Analyzer Service Manual...
Страница 161: ...Electrical Parts List 8 2 TLA600 Series Logic Analyzer Service Manual...
Страница 165: ...Diagrams 9 4 TLA600 Series Logic Analyzer Service Manual...
Страница 176: ...Mechanical Parts List TLA600 Series Logic Analyzer Service Manual 10 11...
Страница 185: ...Mechanical Parts List 10 20 TLA600 Series Logic Analyzer Service Manual...
Страница 186: ......
Страница 187: ......