Adjustments
Input Voltage Gain
WARNING.
Dangerous voltages will be present on the calibration source and connection cables during the following steps.
Verify that the calibrator is in standby mode before making any connection between the calibrator and the instrument.
1.
Connect the generator to the oscilloscope as shown. (See Figure 13.)
Figure 13: Input to 5500A normal output test setup
2.
Press F2 to select the
fi
rst calibration step in the table. (See Table 13.)
Table 13: Input voltage gain calibration points
Cal step
UUT input value (5500A NORMAL)
CL 0799
5 mV
CL 0800
12.5 mV
CL 0801
25 mV
CL 0802
50 mV
CL 0803
125 mV
CL 0804
250 mV
CL 0805
500 mV
CL 0806
1.25 V
CL 0807
2.5 V
CL 0808
5 V
CL 0809
12.5 V
CL 0810
25 V
CL 0811
50 V (set 5500A to OPR!)