Use these performance veri
fi
cation procedures to verify that the oscilloscope meets the warranted speci
fi
cations. These
procedures should be performed in the order given using the equipment recommended in the following table. (See Table 1.)
The values given here are valid for ambient temperatures between 18 °C and 28 °C.
If your oscilloscope fails to qualify under these conditions, service-level adjustments are required. These adjustments
must be done by a trained service person.
WARNING.
The following instructions are for use by quali
fi
ed service personnel only. To avoid electrical shock, do not
disassemble the oscilloscope while it is connected to a signal source other than those speci
fi
ed in this procedure.
Test Equipment
Table 1: Recommended test equipment
Equipment
Minimum requirements
Recommended test equipment
1
Calibration generator
Multiple signal outputs, including
television video signals
2
Fluke 5500A with SC300 or SC600
oscilloscope calibration option, or
Wavetek 9100 and Wavetek 9500 with
active head
Coaxial cable (4)
BNC, 50
Ω
, 36 in
012-0482-xx
Coaxial cable
75
Ω
, 36 in
012-1339-xx
Termination, feedthrough
50
Ω
male-to-female BNC
011-0049-xx
Termination, feedthrough
75
Ω
male-to-female
011-0055-xx
Adapter
BNC female-to-female barrel
103-0028-xx
Adapter (3)
BNC male-to-dual female (T-connector)
103-0030-xx
Adapter
BNC female-to-dual banana plug
103-0090-xx
Voltage probe
10X
THP0301
1
Nine digit part numbers (xxx-xxxx-xx) are Tektronix part numbers.
2
If the SC600 option is not available, a separate television signal generator may be substituted. The generator must be capable of producing video in
the format(s) that you want to test on the THS3000 series oscilloscope; PAL, PALplus, SECAM, and NTSC standards are supported.