Performance Veri
fi
cation
12.
Set the generator to source a 0.8 ms time marker (MODE marker).
13.
Position the trace to the right until the indicated trigger delay is 799.0
μ
s.
Figure 8: 0.8 ms time marker
14.
Examine the rising edge of the time marker pulse at the height of the trigger level indicator top. (You may need to adjust
the horizontal position to view the pulse edge.) Verify that the rising edge is at the second grid line from the left. The
allowed deviation is ±3 pixels. (See Figure 7.)
15.
Set the generator output to 0 V and to standby.
Technical Reference
15