Adjustments
Calibration Procedure
Begin the calibration procedure here.
Check the Instrument Date
1.
Press the Utility button and select OPTIONS.
2.
Use the arrow buttons to select DATE ADJUST and then press the ENTER button.
3.
Adjust the date if necessary and then press the ENTER button.
Select the Calibration Mode
1.
Press and hold the Utility button.
2.
Press and release the Menu Off button.
3.
Release the Utility button.
The CAL MODE (Calibration Adjustment) screen appears and shows:
the calibration step Warming Up (CL 0200)
the calibration status :IDLE (valid) or :IDLE (invalid)
the softkey menu
NOTE.
You can leave the calibration mode without changing the calibration data by turning the instrument off.
Screen messages and key functions.
When the instrument is in the calibration mode, only the four soft keys at the
bottom of the screen, the power key, and the Menu Off key can be operated, unless otherwise stated. The calibration
adjustment screen shows the actual calibration step (name and number) and its status: Cal Name (CL nnnn) :Status (...).
Table 10: Screen message and key function descriptions
Item
Description
Cal Name
Name of the selected calibration step, for example, WarmingUp
(CL nnnn)
Number of the calibration step
Status (...) can be:
IDLE (valid)
After (re)entering this step, the calibration process is not started. The calibration data
of this step are valid. This means that the last time this step was done, the calibration
was successful. It does not necessarily mean that the unit meets the speci
fi
cations
related to this step!
IDLE (invalid)
After (re)entering this step, the calibration process is not started. The calibration data
are invalid. This means that the last time this step was done, the calibration was
not successful. Most probably the unit will not meet the speci
fi
cations if the actual
calibration data are saved.
BUSY aaa% bbb%
Calibration adjustment step in progress; progress % for Input A and Input B. During
WarmingUp the elapsed time is shown.
Technical Reference
25